Dimensions (in) | 8.6 H x 22.8 W x 15 D |
Dimensions (cm) | 22 H x 58 W x 38 D |
Weight | 30 lbs (13.6 kg) |
Power Source | 90-250 VAC, 50/60 Hz |
Performance Specifications |
|
Wavelength Range | 190-1000nm |
Wavelength Selection | Monochromator, tunable 1.0 nm increments |
Wavelength Bandwidth | ≤ 2.0 nm |
Wavelength Accuracy | < ± 1.0 nm |
Wavelength Repeatability | ± 0.2 nm |
Photometric Range | 0.0 to 4.0 nm |
Photometric Resolution | 0.001 OD |
Photometric Accuracy (Microplate) | < ± 0.006 OD ± 1.0%, 0.0 - 2.0 OD |
Photometric Accuracy (Cuvette) | < ± 0.005 OD ± 1.0%, 0.0 - 2.0 OD |
Photometric Precision | < ± 0.003 OD ± 1.0%, 0.0 - 2.0 OD |
Stray Light | < 0.05% @ 230 nm |
Light Source | Xenon Flash Lamp |
Microplate Read Time - Endpoint | 96-Well: 9 seconds 384-Well: 29 Seconds |
Microplate Read Time - Kinetic | 96-Well: 9 second interval 384-Well: 29 second internal |
Microplate Read Time - Speed Read Mode | 96-Well: 5 seconds 384-Well: 16 seconds |
Temperature Range | Ambient + 4 °C up to 45 °C |
Temperature Uniformity (Microplate) | ± 0.5 °C @ 37 °C |
The SpectraMax Plus384 Microplate Reader supports a multitude of absorbance assay categories including:
Product | Part Number |
SpectraMax Plus 384 Microplate Reader | PLUS 384 |
SpectraTest ABS1 Absorbance Validation Plate | 0200-6117 |
StakMax Microplate Handling Stacker | StakMax |
Quartz 96-well microplate for UV measurements | R8024 |
Quartz 96-well half-area microplate for UV measurements | R8028 |
SpectraDrop™ Micro-Volume Microplate Starter Kit | 0200-6262 |
SpectraDrop™ Micro-Volume Microplate HTS Kit | 0200-6267 |
Cuvette Validation Set | 0200-2420 |
Microplate Reader Shelf | 9000-0756 |
SoftMax Pro Software simplifies setting up your assays by providing:
SoftMax Pro Software offers a variety of customizing options:
AI Plácido, CR Oliveira, PI Moreira, CMF Pereira - Molecular neurobiology, 2015 - Springer
... The blue formazan crystals formed were dissolved in an equal volume of 0.04 M HCl in
isopropanol and quantified spectrophotometrically by measuring the absorbance at 570 nm using
a microplater reader (SpectraMax Plus 384, Molecular Devices, San Francisco, CA, USA). ...
Z Huang, Y Huang, H He, J Ni - Molecular medicine …, 2015 - spandidos-publications.com
... described (18). The 96‑well plate used was purchased from Thermo Fisher Scientific
(Beijing, China) and the microplate reader (SpectraMax Plus 384) was purchased
from Molecular Devices, LLC (Beijing, China). The half ...
S Zhang, W Tang, L Jiang, Y Hou, F Yang… - Biotechnology & …, 2015 - Taylor & Francis
... After dilution of the reaction mixture with 3 mL of distilled water and centrifugation, the formation
of cinnamate was determined by measuring the absorbance increase at 290 nm with a SpectraMax
Plus 384 Microplate Reader (Molecular Devices, USA) against a corresponding ...