Molecular Devices microplate readers have the most variety of tools available for users to meet their diverse hardware and software validation needs. Molecular Devices provides complete solutions with:
General Specifications |
|
Dimensions (in.) | 16.5 W x 17 D x 8.75 H |
Dimensions (cm) | 42 W x 43 D x 22 H |
Weight | 36 lbs (16.4 kg) |
Power source | 230 Vac 50 Hz, 115 Vac 60 Hz |
Detector | Low-noise photomultiplier tube with simultaneous photon counting and analog mode |
Plate format | Standard 96- and 384-well microplates, with heights from 13.5-15.5 mm |
Performance Specifications |
|
Cross-talk (96-well) | 3 x 10-5 |
Read modes | Endpoint, Kinetic, Fast Kinetic, Dual-Read |
Injector | 2 variable volume injectors (10-100 µl) per read head |
Injector accuracy | ±1 µL ±2% |
Injector precision | ±1 µL ±2% |
Sensitivity (96-well) | < 20 attomol ATP per well, < 0.2 fg firefly luciferase per well |
Spectral range | 380-630 nm |
Dynamic range | 9 orders of magnitude |
Heater | Temperature range: ambient +5°C to 45°C |
The SpectraMax L Microplate Reader supports a number of flash and glow luminescence assays including:
Product |
Product |
SpectraMax L Microplate Reader | SpectraMax L Config |
BRET1 Filter Set (480nm/535nm) | 0310-5706 |
BRET2 Filter Set (410nm/515nm) | 0310-5707 |
Barcode reader | 0310-5743 |
SpectraTest LM1 Luminescence Validation Plate | 0200-6186 |
StakMax Handling Stacker Base Unit | STAKMAX |
Microplate Reader Shelf | 9000-0756 |
MAE Schoeman, MJC Moester… - Journal of Cellular …, 2015 - Wiley Online Library
... After 24 hours of treatment with the indicated reagents, luciferase activity was determined
using the Dual-Glo Luciferase assay system (Promega, Fitchburg, WI, USA) with a
SpectraMax L luminometer (Molecular Devices, Sunnyvale, CA, USA). ...
FM de Melo, CJM Braga, FV Pereira… - Immunology and cell …, 2015 - nature.com
... The reaction was developed with ECL (EMD Millipore), read in Spectramax L equipment
(Molecular Devices) and expressed as Relative Luminescent Units. For immunoglobulin
isotyping, we used the protocol described by Bargieri et al. ...
V Kozlovskaya, B Xue, W Lei… - Advanced …, 2015 - Wiley Online Library
... of 200 μL. Luminescence was quantified using a SpectraMax L Luminescence
microplate reader and was recorded every 2 min for 1 h after stimulation with 100
ng mL −1 PMA and 1μg mL −1 In. Western Immunoblotting: Whole ...
M Durben, D Schmiedel, M Hofmann, F Vogt… - Molecular …, 2015 - nature.com
... 6 hours. Luminescence induced after activation of the Jurkat cells by bispecific
antibodies bound to the NALM-16 cells was measured in a SpectraMax L luminometer
(Molecular Devices, Biberach, Germany). Serum concentrations ...
KJ Hassett, DJ Vance, NK Jain, N Sahni… - Journal of …, 2015 - Wiley Online Library
... Cell viability was assessed using Cell Titer Glo (Promega, Madison, Wisconsin) and a
Spectramax L Luminometer (Molecular Devices). Neutralizing titers were defined as the
inverse titer that protected at least 50% of the cells from LT. RESULTS. ...
SY Cho, DG Lee, SM Choi, C Park… - BMC infectious …, 2015 - biomedcentral.com
... Relative luminescence units (RLU) were measured using a SpectraMax L luminescence
microplate reader (Molecular Devices, Sunnyvale, CA, USA). The percentage of RLUs compared
to the antibiotic-free controls (%RLU) was calculated, with minimal Page 6. ...
MJ Piña, SM Alex, FJ Arias, M Santos… - … Research Part A, 2015 - Wiley Online Library
... Promega). The Bright Glo luciferase assay system (Promega) was added in a 1:1
proportion of lysate and luciferase reagent. The light produced was measured using
a luminometer (SpectraMax L, Molecular Devices). The protein ...
KM Chung, H Park, S Jung, S Ha, SJ Yoo… - STEM …, 2015 - Wiley Online Library
... A freshly prepared dilution of the Calpain-Glo Reagent stock solution was added to the cells and
luminescence was recorded with a plate-reading luminometer (Molecular Devices, Sunnyvale,
CA, http://www.moleculardevices.com; SpectraMax L). The luminescence values ...
CY Ku, YR Wang, HY Lin, SC Lu, JY Lin - 2015 - dx.plos.org
... 40 min. Finally, 50 μL kinase detection reagent was added to introduce luciferase
and samples were measured with a SpectraMax L Microplate reader (Molecular
Device, CA, USA). Rho GTPase activity assay. Huh7 cells were ...
JSG Muñoz, DJ Rodríguez, JJH Morante - Journal of physiology and …, 2015 - Springer
... levels of biological- ly active forms of GLP-1 (GLP17–36 amide and GLP17– 37) were measured
by fluorescence ELISA through the human glucagon-like peptide 1 (active) kit (Millipore, Billerica,
MA) with a luminescence multi-plate reader (SpectraMax L, Molecular Devices ...
* as of June 26, 2015. Source: Google Scholar.
Forgalmazott termékeink gyártói - keressen gyártó szerint a logóra kattintva