Implement the FDA PAT initiative for on-line release for TOC and Conductivity eliminating time consuming and costly laboratory analysis.
The PAT700 internally captures two samples and then re-analyzes the first sample to validate the excursion alarm and saves the second sample for further root-cause analysis
.
The PAT700 clean-in-place analysis maximizes production up-time by providing immediate on-line results for both TOC and conductivity. Providing instant on-line verification that your production vessel is ready for production changeover.
Using the ASX-7200 auto-sampler allows laboratory users to set up 36 bottles for TOC analysis and walk away. Using the PAT700 there is no requirement for costly reagents or carrier gases.
TOC | Operating Range | 0.5 to 2,000 ppb as Carbon |
Display Resolution | 0.1 ppd | |
Accuracy | +/-1 ppd or +/-5% whichever is greater | |
Repeatability | +/-0.3 ppd or +/-1% whichever is greater | |
Limit of Detection | 0.5 ppb | |
Maximum Input Conductivity | 0.2 uS/cm for all waters, 1.0 uS/cm for all neutral waters, 5.0 uS/cm for water with CO2 as the sole conductive species | |
Conductivity | Conductivity | Range 0.05 to 150 uS/cm (@25* C) |
Display Resolution | 0.01 uS/cm | |
Conductivity Accuracy | +/-2% over full range (uncompensated) | |
Resistivity | Resistivity Range 0.2 to 18 MQ-cm (@25*C) | |
Display Resolution | 0.01 over full range | |
Available Modes | Temperature compensated to 25*C, or uncompensated | |
Temperature | Ambient Operating Range | 10 to 40*C (50 to 104*F) |
Measurement Accuracy | +/-0.4*C | |
Sample Water Range | 5 to 95*C (41 to 203*F) | |
Display Resolution | 0.1 over full range | |
Physical Specs | UV Lamps | 2, with UV Detect technology |
Interface/Display | Color with touch screen | |
Maximum Altitude | 4,000 m (13,125) | |
User I/O Wiring | Three, 3/4-inch conduit openings or quick disconnect fittings | |
Standards System | Onboard, Automated Standards Introduction System (OASIS) | |
Dimensions | 59.7 w X 22.9 d X 25.4 h (23.5 X 9 X 10 inches) | |
Weight | 13.6 kg (30 lbs) | |
Sample Inlet Flow Rate Range | 60 mL/min to 300 mL/min | |
Sample Inlet Pressure Range | 10 to 100 psi (69 to 690 kPa) | |
Compliance | Installation Category | II |
Pollution Degree | 2, IEC 61010-1 | |
CE Compliance | EN 61010-1 and EN 61326 | |
Safety Rating | ETL, conforming to UL 61010-1 and CSA 22.2 No. 61010-1 | |
Enclosure Rating | Conduit version: IP56 Quick connect version: IP46 | |
Release Tests | USP , USP , JP 16, EP 2.2.44 | |
New Features | CIP | Selectable mode for Clean-In-Place analysis |
Dual Stream option | Toggle or programmable stream switching | |
Excursion sampling | Minimum flow rate to fill excursion bottle = 160 mL/min | |
Rouge detection | Identifies oxidation cell contamination from roughing |
Physical Specs | Vial Capacity | 36 Grab sample bottles HDPE or Borossilicate USP Type 1 glass |
Dimensions | 45.7 cm X 36.6 cm X 53.7 cm (HxWxD) | |
Weight | 21 kg (45 lbs) | |
Communications | Communications | RS232 interface with PAT700 - No external PC required |
Environment | Ambient Environment | 10 - 40*C, 5 - 95% non-condensing RH |
Power | Power Requirements | 100 - 240 VAC +/- 10%, 4.4A |