JM Modern HikaShop - шаблон joomla Видео
In
(06-1) 619-8321
BIO SCIENCE
Nuaire

B2 fülke, NU-430, 100% szennyezett levegő elszívással

nu-430_left_1
nu-430_right_4
The LabGard ES 430 Total Exhaust Biological Safety Cabinet may be used for work with non-flammable or explosive volatile chemicals and radionuclides by providing personnel, product, and environmental protection by exhausting 100% of contaminated air.

ÁRAJÁNLATOT, INFORMÁCIÓT KÉREK!

Exhaust Interlock

The Nu-430 contains an exhaust interlock system that prevents operation of the internal supply blower unless the exhaust flow is sufficient to provide the correct air barrier inflow velocity at start up.

TouchLink2 Electronic Control System

An easy to use color touch screen operates all system functions.

  • On/off for lights, blower, and interior outlets.
  • Monitors high/low limits for down flow, inflow and sliding window position.
  • Password protection locks out unwanted changes to the system.
  • Date/clock display and timer function can be used as a laboratory timer, to set purge cycles, outlet timer, UV light timer, auto-run timer, NiteCare night set back, or weekend turn-off.
  • Filter status displays remaining filter life
  • Complete diagnostic functions for a NSF trained service technician or certifier.
  • NEW – BACnet RS-485 interface provides the Building Automation System (BAS) a two-way communication link to query the TouchLink2 operating information. View airflow data, alarm status, etc. and control TouchLink2 for facility energy savings by initiating NiteCare (Night Setback System).

IntelliFlow Airflow Control

Assure optimal biological safety cabinet performance by monitoring, displaying, and controlling real time airflow with the same methods of you annual certifier with a dual thermistor airflow sensors powered by measurement instrument company TSI™ controlling downflow and a pressure flow grid controlling exhaust flow.

NiteCare

Night setback system allows for an energy saving sterile working environment when the cabinet is not in use.

Monolithic Pressure Tight Construction

A single shell, high grade stainless steel, fully welded cabinet eliminates potential leaks that may occur through failures of gaskets and cabinet seams.

Energy Saver Motor

NuAire ultra-high efficiency motor technology lowers energy costs while increasing filter life from the NSF minimum requirement of 50% (3 year equivalent life) to 300% or higher (10 or more years).

Experience Ergonomics – Over 640 muscles at work.

NuAire Biological Safety Cabinets are designed to create an ergonomically correct environment reducing user fatigue and work place injuries.  

  • Maximum knee/thigh clearance that improves the user’s ability to assume a proper posture. The maximum knee/thigh clearance is accomplished by a design improvement that reduced the plenum height beneath the work surface.
  • An adjustable base stand provides the ability for laboratory personnel to optimize ergonomic conditions improving leg and forearm support. A heavy duty plastic armrest minimizes cold conduction and provides additional forearm support.
  • Cool white lighting and a polished edge window reduces glare allowing for greater visibility and better sight lines to the effective work zone area reducing awkward posture resulting in less eye fatigue and/or neck strain.
  • NuAire Biological Safety Cabinets provide the largest effective work zone area available which helps reduce arm/neck/shoulder strain.
  • Our ultra-high efficiency motor runs quieter with fewer vibrations creating less stress on the body.
  • Additional accessories such as an adjustable footrest, elbow rests, turntable, and ergonomically designed chair enhances the user’s ergonomic working environment.

TruLaminar Airflow

Unidirectional airflow moving along parallel flow lines at a constant velocity of 60 fpm (0.30 m/s) minimizes air turbulence within the work zone lessening the chance for cross contamination.

Dynamic Air Barrier

Strong inflow air velocity of 105 fpm (0.53 m/s) over an aerodynamic airfoil creates maximum personnel protection.

HEPEX™ Zero Leak Airflow System

The HEPEX™ plenum inflates to provide quiet uniform air without transferring vibration. Positive pressure chambers and ducts are surrounded by negative pressure eliminating potential leaks. HEPEX™ fabrics are incredibly durable. They are flameproof, provide a barrier against water, gases, fluid, and dust as well as resist to chemicals, oils, and greases. It even holds antibacterial properties and is impregnable meaning harmful materials become trapped in the filters.

Attenumount

Vibration and sound control

Biological Test Assurance

NuAire biosafety cabinets are tested in laboratory extremes to ensure maximum personnel, product, and environmental protection.

Standard Features

Motor speed control

Internal exhaust dampener

Permanent Plenum w/ Quick Release

Supply HEPA Filter 99.99% @ 0.3 microns

Exhaust HEPA Filter 99.99% @ 0.3 microns

Front Filter Removal

Fluorescent Cool White Lighting

Frameless Sliding Tempered Glass

Heavy Duty Ergo Armrest

Removable Work Tray

Metal Diffuser over Supply Filter

Duplex Outlets

Services Valves

Optional Features

Germicidal UV Light w/ Window Interlock

Manual Stainless Steel Butterfly Valve

Automatic Stainless Steel Butterfly Valve

Bag-In Bag-Out Filter Exchange System

GFI

Additional Duplex Outlets

IV Bar w/ Hooks

Base Support Stands

Storage Support Stands

Adjustable Support Stands

Hinged Viewing Window

Microscope Window

Additional Service Valves

Remote Service Valves

Metal Framed HEPA Filters 99.99% @ 0.3 Microns

ULPA Filters 99.999% @ 0.12 Microns

Sorbent Exhaust Filter Module

Prefilter for Supply Air

Storage Pull-Out Trays

Decorative Side Panels

Sinks w/ Hot/Cold or DI Faucets

Heated Work Surface

Seismic Brackets

Custom Solutions

 

Műszaki specifikációk
Model NU-430-300 NU-430-400 NU-430-600
Nominal Size 3 ft. (0.9 m) 4 ft. (1.2 m) 6 ft. (1.8 m)
Style of Cabinet Bench top/console with optional base stand Bench top/console with optional base stand Bench top/console with optional base stand
Sash Height 8" (203 mm) 8" (203 mm) 8" (203 mm)
Exterior Dimensions (W x D x H)
in. [mm]
37 ⅝ x 32 ⅞ x 61 
(1057 x 835 x 1549)
53 ⅝ x 32 ⅞ x 61 
(1362 x 835 x 1549)
77 ⅝ x 32 ⅞ x 61
(1972 x 835 x 1549)
Interior Dimensions (W x D x H)
in. [mm]
30 ⅜ x 23 ½ x 25 ½ 
(772 x 597 x 648)
46 ⅜ x 23 ½ x 25 ½ 
(1178 x 597 x 648)
70 ⅜ x 23 ½ x 25 ½ 
(1788 x 597 x 648)
Weight 460 lbs. / 231 kg. 520 lbs. / 236 kg. 710 lbs. / 322 kg.
Inflow Velocity 105 fpm (.53 m/s) 105 fpm (.53 m/s) 105 fpm (.53 m/s)
Downflow Velocity 60 fpm (0.30 m/s) 60 fpm (0.30 m/s) 60 fpm (0.30 m/s)
Heat Rejected, BTU, Per Hour
(non-vented)
(vented)
     
Sound Pressure Level  
Light Intensity foot-candles (LUX) 80 to 150 (860 to 1600) 80 to 150 (860 to 1600) 80 to 150 (860 to 1600)
Accolades NSF/ANSI 49, EN12469, ULUL-C, CE, Energy Saver, Personnel Product Environmental Protection, RoHS
Applikációk
  • Learn how to work safely in your NuAire Biological Safety Cabinet
  • Understanding the NU-430 / NU-435 Class II, Type B2 Biosafety Cabinet / Fume Hood
  • CETA Application Guide for the Exhaust System Requirements of Class II, Type B Biosafety Cabinets
  • Optional Lay-in Charcoal Filter Module
  • Air filtration and the use of HEPA filters in Biosafety Cabinets
Videók, ismertetők

blue down arrowForgalmazott termékeink gyártói - keressen gyártó szerint a logóra kattintva