SoftMax® Pro Data Acquisition & Analysis Software, Molecular Devices industry leading all-in-one data acquisition and analysis software, offers:
General Specifiations |
|||||||||||||||||||||||||||||
Dimensions (in.) |
8.6 (H) x 22.8 (W) x 15 (D) |
||||||||||||||||||||||||||||
Dimensions (cm) |
22 (H) x 58 (W) x 38 (D) |
||||||||||||||||||||||||||||
Weight |
35 lbs. (15.9 kg) |
||||||||||||||||||||||||||||
Power consumption |
< 125 watts |
||||||||||||||||||||||||||||
Power source |
100-240 Vac, 3 A, 50/60 Hz |
||||||||||||||||||||||||||||
Absorbance Photometric Performance |
|||||||||||||||||||||||||||||
Wavelength range |
200-1000 nm |
||||||||||||||||||||||||||||
Wavelength selection |
Monochromator, tunable in 1.0 nm increments |
||||||||||||||||||||||||||||
Wavelength bandwidth |
≤ 4.0 nm |
||||||||||||||||||||||||||||
Wavelength accuracy |
±2.0 nm |
||||||||||||||||||||||||||||
Wavelength repeatability |
±0.2 nm |
||||||||||||||||||||||||||||
Photometric range |
0-4.0 OD |
||||||||||||||||||||||||||||
Photometric resolution |
0.001 OD |
||||||||||||||||||||||||||||
Photometric accuracy (microplate) |
< ±0.006 OD ±1.0%, 0-2 OD |
||||||||||||||||||||||||||||
Photometric accuracy (cuvette) |
< ±0.005 OD ±1.0%, 0-2 OD |
||||||||||||||||||||||||||||
Photometric precision |
< ±0.003 OD ±1.0%, 0-2 OD |
||||||||||||||||||||||||||||
Baseline flatness |
< 0.001 OD |
||||||||||||||||||||||||||||
Stray light |
< 0.05% @ 230 nm |
||||||||||||||||||||||||||||
Fluorescence Photometric Performance (M2) |
|||||||||||||||||||||||||||||
Dual monochromators |
1 nm increments |
||||||||||||||||||||||||||||
EX |
250-850 nm |
||||||||||||||||||||||||||||
EM |
250-850 nm |
||||||||||||||||||||||||||||
Bandwidth (EX, EM) |
9, 9 nm |
||||||||||||||||||||||||||||
Optimized sensitivity (top-read) |
5 pM (fluorescein) |
||||||||||||||||||||||||||||
Guaranteed sensitivity (top-read)** |
15 pM (fluorescein) |
||||||||||||||||||||||||||||
Fluorescence Ph
Applikációk
Molecular Devices Assay Partnerships Molecular Devices has collaborated with various assay partners to optimize and validate homogeneous and heterogeneous biochemical- or cell-based assay performance on the SpectraMax Microplate Reader platform. To support these assays, we provide application notes as well as ready-to-run protocols in our SoftMax Pro Software. The SpectraMax microplate readers support a wide variety of assay categories including
Tartozékok, kiszerelések
SpectraMax® M Series Multi-Mode Microplate Readers
Videók, ismertetők
Related Video
Related Collateral
Applications
Poster
Number of Citations*: 11,100
Quantification of Catecholamine Uptake in Adult Cardiac MyocytesEF Dahl, CD Wright, TD O'Connell - Nuclear G-Protein Coupled Receptors, 2015 - Springer ... 1. FlexStation 3 Benchtop Multi-Mode Microplate Reader. 2. SpectraMax M5 Multi-Mode Microplate
[HTML] Determining Antioxidant Activities of Lactobacilli Cell-Free Supernatants by Cellular Antioxidant Assay: A Comparison with Traditional MethodsJ Xing, G Wang, Q Zhang, X Liu, Z Gu, H Zhang… - PloS one, 2015 - dx.plos.org ... The control sample contained deionized water instead of the sample solution. The
Forgalmazott termékeink gyártói - keressen gyártó szerint a logóra kattintva |