Detector |
Photomultiplier tubes (PMT) and/or photodiode |
2 PMT |
Photo-multiplier tube |
2 PMT |
1, 2, or 6 Ultra-fast Photomultiplier Tubes |
Spectral range |
300-850 nm |
300-850 nm |
250-850 nm |
250-850 nm |
380-630 nm |
Dynamic range |
> 6 decades |
> 6 decades |
> 6 decades |
> 6 decades |
> 9 decades |
Sensitivity |
15 pM for 96-wells, 30 pM for 384-wells (Optimized for ATP-Glow) |
1.25 pM for 96-wells, 3 pM for 384-wells, 19 pM for 1536-wells (Optimized for ATP-Glow) |
43 pM ATP in 96 wells (optimized) |
< 20 attomol ATP per well |
Microplate format |
6 to 1536 wells* |
6- to 1536-wells |
6-384-well microplates |
6-384 (96- and 384-wells with pipetting) |
96- and 384-well microplates |
Cross-talk |
< 0.3% in white 96- and 384-well microplates |
|
<0.3% in white 96- and 384-well microplates |
<0.3% in white 96- and 384-well microplates |
< 3 x 10-5 (96-well) |
Read modes |
Spectral, Endpoint, Kinetic |
Endpoint, Kinetic |
Endpoint, Kinetic |
Endpoint, Kinetic, Fast Kinetic |
Endpoint, Kinetic, Fast Kinetic, Dual-read |
Read times |
0:30 minutes for 96-wells, 1:15 minutes for 384-wells |
|
2:00 minutes for 96-wells, 7:00 minutes for 384-wells |
2:00 minutes for 96-wells, 7:00 minutes for 384-wells |
0.1 up to 3600 sec. in steps of 0.1 sec. |
Injectors/pipettors |
Option available |
Option available |
|
8- or 16-channel pipettor |
2 variable volume injectors |
Injection/pipettor volumes |
|
|
|
Max. volume 200uL for 8-channel and 30 uL for 16-channel |
|
Temperature control |
4 degrees above ambient to 45 degrees C |
2 degrees C above ambient to 45 degrees C |
2 degrees C above ambient to 60 degrees C |
2 degrees C above ambient to 45 degrees C |
5 degrees C above ambient to 45 degrees C |
Software |
SoftMax Pro Software |
SoftMax Pro Software |
SoftMax Pro Software |
SoftMax Pro Software |
SoftMax Pro Software |
Validation & compliance tools |
IQ/OQ Protocols, SoftMax Pro GxP Compliance Software, SpectraTest LM1 Validation Plate |
SoftMax Pro GxP Compliance Software |
IQ/OQ Protocols, FDA 21 CFR Part 11 Compliance, SpectraTest LM1 Validation Plate |
IQ/OQ/PQ Protocols, FDA 21 CFR Part 11 Compliance |
Software Validation Package, IQ/OQ Documentation, Physical Validation Test Plate |
Robotic compatible |
x |
x |
x |
x |
x |