MET ONE R4800 series remote air particle counters use standard RS-485 serial communications MET ONE R4900 series remote air particle counters use industry standard 4-20mA interface
Sensitivity ranges from 0.3um to 0.5um and flow rates ranging from 0.1cfm to 1.0cfm Onboard LED status indicator that flashes and sets a signal at the connector for alarm Stay resident RS-485 configuration programming (count time, hold time, alarm level, alarm delay and scaling) is maintained through power cycles (power on/off)
The small stainless steel enclosure is compatible with most process environments. There is no fan to disturb airflows, allowing the MET ONE R4800/R4900 series counters to be placed close to sensitive process areas.
Minimum threshold Setting | R4803 / R4903 - 0.3um R4805 / R4815/ R4905 / R4915 - 0.5um |
Flow Rate | R4803 / R4805 / R4903 / R4905 - 0.1 CFM (2.83 LPM) R4815 / R4915 - 1.0 CFM (28.3 LPM) |
Flow Control | Critical orifice; requires 18” Hg vacuum, minimum 450 mbar |
Size Channels (μm) | R4803 / R4903 - Ch 1 / Ch2 - 0.3 / 0.5 R4805 / R4815 / R4905 / R4915 - Ch 1 / Ch2 - 0.5 / 5.0 |
Light Source | Laser diode (10-year MttF) |
Coincidence Loss | 5% at 2,000,000 particles/ft3 |
Inlet Pressure | Ambient to 0.1” Hg vacuum |
Indicators | Power LED illuminates when counter is powered; Calibration/Alarms LED illuminates when sensor calibration needs checking and flashes when the count exceeds the alarm level. |
Power | Open collector Fet maximum sink current (on) 150 mA; Maximum VC voltage (off) 30 VDC |
Connector | DB-15 (female) |
Weight | 0.3 kg (10.7 oz) |
Tubing Connections | Inlet, 0.1 CFM, 1/8” I.D. Inlet, 1.0 CFM, 1/4” I.D. Outlet, All (vacuum), 1/4” I.D. |
Environment | Operating - 12to 41C (55 to 105F) 20 to 95% relative humidity, non-condensing Storage - -40 to 70C (-40 to 160F) up to 98% relative humidity, non-condensing |
When ordering, specify | 0.3 µm (R4803/R4903) or 0.5 µm (R4805/R4815/R4905/R4915) AND 0.1 cfm (R4803/R4805/R4903/R4905) or 1.0 cfm (R4815/R4915) |
Included Accessories
Part Number | Description |
---|---|
500052 | Isokinetic probe; DB-15 Connector |
701113 | Operators Manual R4800 |
701130 | Operators Manual R4900 |
Optional Accessories
Part Number | Description |
---|---|
2080442 | Purge Filter Assembly, 0.2µ , 0.1 cfm (R4803 / R4805 / R4903 / R4905) |
2080443 | Purge Filter Assembly, 0.2µ , 1.0 cfm (R4815/R4915) |
2082270-1 | Power Supply, 110 VAC input, 12 vdc output supports up to 32 model R4800 counters |
2082270-2 | Power Supply, 230 VAC input, 12 vdc output supports up to 32 model R4800 counters |
2081524-11 | Wall Plate, R4800/R4900 Series |